FDA Device recall Z-2961-2020

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sa…

On 2020-09-23 the FDA classified a Class II recall of Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sa… by Terumo Cardiovascular Systems, citing internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2961-2020
ClassificationClass II
Statusterminated
Initiated2020-08-25
Published2020-09-23
Last updated2026-09-13 11:55 UTC
CompanyTerumo Cardiovascular Systems
DistributionUS

Product

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.

Reason

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Identifiers

code_infoUDI: 00886799000052; Serial Numbers: 20019 20020

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.