FDA Device recall Z-2961-2020
Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sa…
- FDA Device
- Class II
- terminated
- Published 2020-09-23
On 2020-09-23 the FDA classified a Class II recall of Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sa… by Terumo Cardiovascular Systems, citing internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2961-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-08-25 |
| Published | 2020-09-23 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.
Reason
Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.
Identifiers
| code_info | UDI: 00886799000052; Serial Numbers: 20019 20020 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2961-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.