FDA Device recall Z-2960-2026

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

On 2026-08-26 the FDA classified a Class II recall of TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799 by Tobii Dynavox, citing under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit.….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2960-2026
ClassificationClass II
Statusongoing
Initiated2026-07-08
Published2026-08-26
Last updated2026-09-13 11:58 UTC
CompanyTobii Dynavox
DistributionUS

Product

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Reason

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Identifiers

code_info
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is…UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).