RecallRadar

FDA Device recall Z-2957-2024

Impella CP Pump 371 Set, Clinical Trial

On 2024-09-11 the FDA classified a Class II recall of Impella CP Pump 371 Set, Clinical Trial by Abiomed, citing A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2957-2024
ClassificationClass II
Statusongoing
Initiated2024-08-05
Published2024-09-11
Last updated2026-09-13 11:57 UTC
CompanyAbiomed
DistributionUS

Product

Impella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;

Reason

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

Identifiers

code_info
Product Code: 0048-0040; UDI-DI: 00813502011739; Serial Numbers: 370089 370091 370090 370074 370093 401804 401805 40181…Product Code: 0048-0040; UDI-DI: 00813502011739; Serial Numbers: 370089 370091 370090 370074 370093 401804 401805 401811 401806; Batch Numbers: 2023172492 2023172497 2023172494 2023172500 2023172499 2023255511 2023255513 2023255522 2023255514

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2957-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.