Impella CP Pump Set, Japan

On 2024-09-11 the FDA classified a Class II recall of Impella CP Pump Set, Japan by Abiomed, citing A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2956-2024
ClassificationClass II
Statusongoing
Initiated2024-08-05
Published2024-09-11
Last updated2026-09-13 11:57 UTC
CompanyAbiomed
DistributionUS

Product

Impella CP Pump Set, Japan; Product Code: 0048-0034-JP;

Reason

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

Identifiers

code_infoProduct Code: 0048-0034-JP; UDI-DI: 00813502011463; Serial Numbers: 324303 323153 328083 328085; Batch Numbers: 2022079852 2022079846 2022083745 2022083747