FDA Device recall Z-2943-2020
Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - P…
- FDA Device
- Class II
- terminated
- Published 2020-09-16
On 2020-09-16 the FDA classified a Class II recall of Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - P… by Talladium, citing the linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2943-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-08-06 |
| Published | 2020-09-16 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Talladium |
| Distribution | US |
Product
Talladium Luminesse Zirconia Discs, Zirconia HT Disc 98x20mm A1, Shaded Zirconia Blend 2.0 50# A1, Model No. 5147-A1;(UDI): D7915147A10 - Product Usage: are intended for CAD/CAM fabrication of all-ceramic (no metal) dental restorations. Luminesse ZR Blanks are partially sintered Yttria stabilized zirconia blanks for use as CAD/CAM milling blanks.
Reason
The linear shrinkage on the zirconia blank disc s part label is incorrect. The average linear shrinkage found on the product label was outside the acceptable range. The acceptable average linear shrinkage ranges between 1.225 to 1.275. The production lot being recalled has a linear shrinkage rate of 2.251 which is above the acceptable range.
Identifiers
| code_info | Lot Number.: 1AZG |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2943-2020%22