FDA Device recall Z-2939-2020

BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 4900043…

On 2020-09-16 the FDA classified a Class I recall of BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 4900043… by Carefusion 303, citing pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2939-2020
ClassificationClass I
Statusongoing
Initiated2020-08-04
Published2020-09-16
Last updated2026-09-13 11:55 UTC
CompanyCarefusion 303
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.

Reason

Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.

Identifiers

code_infoAll serial numbers