RecallRadar

FDA Device recall Z-2938-2026

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

On 2026-08-19 the FDA classified a Class II recall of ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000 by Erbe Usa, citing A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2938-2026
ClassificationClass II
Statusongoing
Initiated2026-06-26
Published2026-08-19
Last updated2026-09-13 11:58 UTC
CompanyErbe Usa
DistributionUS

Product

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Reason

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

Identifiers

code_infoUDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.