FDA Device recall Z-2936-2018

Duo-Flow Catheter Full Tray

On 2018-09-05 the FDA classified a Class II recall of Duo-Flow Catheter Full Tray by Medical Components, citing the suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 b….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2936-2018
ClassificationClass II
Statusterminated
Initiated2018-07-12
Published2018-09-05
Last updated2026-09-13 11:54 UTC
CompanyMedical Components
DistributionUS

Product

Duo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis catheter insertion kits

Reason

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

Identifiers

code_info
MKAZ210 MCBK870 MCBS910 MCBZ450 MCCP990 MCCZ440 MCDNlOO MCDQ930 MCFD710 MCFN970 MCFS730 MCAS340 MCBA830 MCBR880 MCBY910…MKAZ210 MCBK870 MCBS910 MCBZ450 MCCP990 MCCZ440 MCDNlOO MCDQ930 MCFD710 MCFN970 MCFS730 MCAS340 MCBA830 MCBR880 MCBY910 MCCF130 MCCX150 MCDF010 MCDG980 MCDP120 MCFD720 MCFJ060 MCFM520 MCFW300 MCFZ590 MKAJ380 MCAS230 MCBL150 MCBN500 MCBZ380 MCCV010 MCDF030 MCFX080 MCFV620 MCBA170 MCDC940 MKAX240 MCAC870 MCCQ000 MCFG380 MKAQ920 MCAH920 MCCW950

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2936-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.