RecallRadar

FDA Device recall Z-2933-2018

Double Lumen CRRT/ABP Catheter Set

On 2018-09-05 the FDA classified a Class II recall of Double Lumen CRRT/ABP Catheter Set by Medical Components, citing the suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 b….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2933-2018
ClassificationClass II
Statusterminated
Initiated2018-07-12
Published2018-09-05
Last updated2026-09-13 11:54 UTC
CompanyMedical Components
DistributionUS

Product

Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15

Reason

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

Identifiers

code_info
MCAJ510 MCAP860 MCAZ820 MCCB890 MCCJ250 MCDF160 MCDQ630 MCFF200 MCFF620 MCFK200 MCFV440 MKAK310 MCAW770 MCBW430 MCCH3…MCAJ510 MCAP860 MCAZ820 MCCB890 MCCJ250 MCDF160 MCDQ630 MCFF200 MCFF620 MCFK200 MCFV440 MKAK310 MCAW770 MCBW430 MCCH320 MCCJ690 MCFB450 MCFF930 MCFV450 MKAK320 MKAV210 MCAY080 MCBD600 MCBP740 MCBW440 MCCA030 MCCH330

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2933-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.