RecallRadar

FDA Device recall Z-2904-2026

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In s…

On 2026-08-12 the FDA classified a Class II recall of Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In s… by Tissium, citing currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2904-2026
ClassificationClass II
Statusongoing
Initiated2026-06-15
Published2026-08-12
Last updated2026-09-13 11:58 UTC
CompanyTissium
DistributionUS

Product

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Reason

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Identifiers

code_infoLot Code: Model No NCOCO2075; UDI-DI ¿03760279380100; Lot Number 250700016, 251000048, 251000049, and 260300043