Philips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare pr…

On 2018-08-29 the FDA classified a Class II recall of Philips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare pr… by Philips Electronics North America, citing philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries shoul….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2896-2018
ClassificationClass II
Statusongoing
Initiated2018-07-13
Published2018-08-29
Last updated2026-09-13 11:54 UTC
CompanyPhilips Electronics North America
DistributionUS

Product

Philips SureSigns VS2 monitor, 863079 863080 863081 863082 863283 863286 The Sure Signs VS2 vital signs monitor is for use by healthcare professionals whenever there is a need for monitoring the physiological parameter of patients. Standard and optional parameters include: NBP, Sp02, Temperature. The Sure Signs VS2 vital signs monitor is for the monitoring, recording and alarming of multiple parameters of adult, pediatric and neonates in healthcare environments. Additionally, this monitor is intended for transport situations within the healthcare facility.

Reason

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

Identifiers

code_infoUDI# VS2 (863079): 00884838009639, (863080): 00884838009646, (863081): 00884838009653, (863082): 00884838009660

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2896-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.