FDA Device recall Z-2888-2026

REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCA…

On 2026-08-12 the FDA classified a Class II recall of REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCA… by Beckman Coulter, citing analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications.".

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2888-2026
ClassificationClass II
Statusongoing
Initiated2026-06-26
Published2026-08-12
Last updated2026-09-13 11:58 UTC
CompanyBeckman Coulter
DistributionUS

Product

REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument

Reason

Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

Identifiers

code_infoUDI: 15099590799243/ V5.03 Panel Update 11

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.