FDA Device recall Z-2878-2018
WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523
- FDA Device
- Class I
- terminated
- Published 2018-09-05
On 2018-09-05 the FDA classified a Class I recall of WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523 by Biosense Webster, citing when attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2878-2018 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2018-05-07 |
| Published | 2018-09-05 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Biosense Webster |
| Distribution | n/a |
Product
WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.
Reason
When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.
Identifiers
| code_info | All lots. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.