FDA Device recall Z-2876-2026
GE Healthcare MAC VU360, REF# part number 2030361-023
- FDA Device
- Class II
- ongoing
- Published 2026-08-12
On 2026-08-12 the FDA classified a Class II recall of GE Healthcare MAC VU360, REF# part number 2030361-023 by Ge Medical Systems Information Technologies, citing software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2876-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-06-29 |
| Published | 2026-08-12 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Ge Medical Systems Information Technologies |
| Distribution | US |
Product
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Reason
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Identifiers
| code_info | GTIN 00198953107263, Serial Numbers: VU221150078SA, VU222400121SA, VU223230048SA, VU222520132SA, VU219400033SA, VU2194…GTIN 00198953107263, Serial Numbers: VU221150078SA, VU222400121SA, VU223230048SA, VU222520132SA, VU219400033SA, VU219400034SA, VU219400045SA, VU219400055SA, VU220270027SA, VU220510048SA, VU220510078SA, VU221100161SA, VU221100181SA, VU221150116SA, VU222120212SA, VU222120220SA, VU222120239SA, VU222120243SA, VU222120249SA, VU222120250SA, VU222120253SA, VU222120258SA, VU222120260SA, VU222120261SA, VU222120266SA, VU222120269SA, VU222140004SA, VU222200296SA, VU222520114SA, VU223120014SA, VU223120041SA, VU223420119SA, VU224050158SA, VU224070091SA, VU224200028SA, VU224420159SA, VU224520140SA, VU225170260SA, VU225420050SA, VU225520037SA, VU226040025SA, VU419320042SA, VU421490135SA, VU421490151SA, VU421490153SA, VU421490155SA, VU421490159SA, VU422460150SA, VU422460194SA, VU422460203SA, VU422460303SA, VU422480164SA, VU422500239SA, VU422500240SA, VU422500241SA, VU422500242SA, VU422500244SA, VU422500245SA, VU422500246SA, VU422500247SA, VU422500248SA, VU422500249SA, VU422500250SA, VU422500251SA, VU |
|---|