RecallRadar

FDA Device recall Z-2876-2018

Persona Partial Knee System Spacer Block Alignment Tower, Item No. 42-5399-006-35 Product Usage: Intended to aid in preparatory surgical pr…

On 2018-08-29 the FDA classified a Class II recall of Persona Partial Knee System Spacer Block Alignment Tower, Item No. 42-5399-006-35 Product Usage: Intended to aid in preparatory surgical pr… by Zimmer Biomet, citing the PPK System Spacer Block Alignment Tower mating feature that mates with the spacer block was undersized, which could potentially lead to mating issues between the devices. This….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2876-2018
ClassificationClass II
Statusterminated
Initiated2018-06-22
Published2018-08-29
Last updated2026-09-13 11:54 UTC
CompanyZimmer Biomet
DistributionUS

Product

Persona Partial Knee System Spacer Block Alignment Tower, Item No. 42-5399-006-35 Product Usage: Intended to aid in preparatory surgical procedures

Reason

The PPK System Spacer Block Alignment Tower mating feature that mates with the spacer block was undersized, which could potentially lead to mating issues between the devices. This could cause a delay in surgery of less than 30 minutes. There is an option in the surgical technique for performing limb alignment without the alignment tower.

Identifiers

code_info63465803, 63762260, 63767434, 63799794

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2876-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.