RecallRadar

FDA Device recall Z-2863-2017

Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in the treatment, incl…

On 2017-08-09 the FDA classified a Class II recall of Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in the treatment, incl… by Spectranetics, citing potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or disengaged rom the catheter when the device….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2863-2017
ClassificationClass II
Statusterminated
Initiated2017-06-06
Published2017-08-09
Last updated2026-09-13 11:54 UTC
CompanySpectranetics
DistributionAK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV

Product

Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in the treatment, including atherectomy, of infrainguinal stenosis and occlusions

Reason

Potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or disengaged rom the catheter when the device is in use in an off-label manner.

Identifiers

code_infoModel Number 414-159 Catheter Diameter 1.4mm, Model Number 417-156 Catheter Diameter 1.7mm, and Model Number 420-159 Catheter Diameter 2.0mm.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2863-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.