FDA Device recall Z-2828-2020

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower ja…

On 2020-08-26 the FDA classified a Class II recall of NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower ja… by Nobel Biocare Services Ag, citing dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lea….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2828-2020
ClassificationClass II
Statusterminated
Initiated2020-07-20
Published2020-08-26
Last updated2026-09-13 11:55 UTC
CompanyNobel Biocare Services Ag
DistributionUS

Product

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Reason

Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.

Identifiers

code_infoLot: 13084072, UDI: (01)07332747073584(10)13084072(17)240525