FDA Device recall Z-2824-2018

Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator

On 2018-08-22 the FDA classified a Class II recall of Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator by Sorin Crm Sas, citing hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrill….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2824-2018
ClassificationClass II
Statusterminated
Initiated2018-07-09
Published2018-08-22
Last updated2026-09-13 11:54 UTC
CompanySorin Crm Sas
DistributionAZ, FL, IL, LA, MI, MS, NJ, OK, PA

Product

Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator

Reason

Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.

Identifiers

code_info
Model TDF033U, UDI GTIN 08031527015538 Serial Numbers: 619DK0FB, 618DK003, 619DK0C0, 622DK03B, 622DK12A, 624D…Model TDF033U, UDI GTIN 08031527015538 Serial Numbers: 619DK0FB, 618DK003, 619DK0C0, 622DK03B, 622DK12A, 624DK070, 624DK12F, 619DK021, 627DK0B0, 618DK0A4, 619DK097, 620DK065, 619DK09B, 619DK05F, 619DK088, 619DK0CC