FDA Device recall Z-2816-2020
DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomie…
- FDA Device
- Class II
- terminated
- Published 2020-08-26
On 2020-08-26 the FDA classified a Class II recall of DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomie… by Synthes Usa Products, citing may have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2816-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-07-21 |
| Published | 2020-08-26 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Synthes Usa Products |
| Distribution | US |
Product
DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for fixation of fractures, osteotomies and non-unions of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula, particularly in osteopenic bone. Part Number 241.381
Reason
May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw
Identifiers
| code_info | Lot Number: 29P6036 UDI: 10886982166913 |
|---|