FDA Device recall Z-2811-2020

Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monit…

On 2020-08-26 the FDA classified a Class II recall of Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monit… by Philips North America, citing A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a c….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2811-2020
ClassificationClass II
Statusterminated
Initiated2020-06-18
Published2020-08-26
Last updated2026-09-13 11:55 UTC
CompanyPhilips North America
Distributionn/a

Product

Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

Reason

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Identifiers

code_infoProduct #: 989803191021 Manufacturing date codes: prior to 2020-03