FDA Device recall Z-2811-2015
Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cra…
- FDA Device
- Class II
- terminated
- Published 2015-09-30
On 2015-09-30 the FDA classified a Class II recall of Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cra… by Synthes Usa Products, citing the Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to desig….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2811-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-08-31 |
| Published | 2015-09-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Synthes Usa Products |
| Distribution | US |
Product
Synthes Cranial Flap Tube Clamp and Crimping Device Product Usage: The Synthes Cranial Flap Tube Clamp System is intended to reattach a cranial bone flap to the surrounding cranium after a craniotomy procedure.
Reason
The Synthes Cranial Flap Tube Clamp and Crimping Device for Cranial Tube Clamp had been labeled MR Safe although they do not have the testing protocols currently required to designate them as MR Safe.
Identifiers
| code_info | Part Number 329.315 460.001 460.001.01S 460.002 460.002.01S 460.003 460.003.01S 460.008 460.008.01S 460.009…Part Number 329.315 460.001 460.001.01S 460.002 460.002.01S 460.003 460.003.01S 460.008 460.008.01S 460.009 460.009.01S 460.010 460.010.01S 460.100 460.100.01S 460.107 460.107.01S with Lot Numbers 4146716, 4154144, 4206682, 4206683, 4213557, 4214824, 4216603, 4216604, 4240689, 4240690, 4241741, 4248462, 4249395, 4249396, 4264386, 4278302, 4278304, 4283035, 4283036, 4283037, 4283038, 4283039, 4283040, 4283041, 4285956, 4289103, 4295084, 4295085, 4298044, 4303093, 4326975, 4326979, 4326981, 4329355, 4331263, 4331265, 4331266, 4344659, 4344678, 4378973, 4482201, 4485317, 4488093, 4488094, 4490608, 4490609, 4496716, 4496915, 4496916, 4501971, 4504043, 4508068, 4513458, 4513459, 4513460, 4513461, 4516371, 4519406, 4519407, 4521408, 4525905, 4550062, 4550063, 4550064, 4550065, 4551164, 4551165, 4551166, 4561400, 4561401, 4572176, 4574481, 4580884, 4586704, 4605094, 4661414, 4669482, 4669483, 4675725, 4695343, 4705970, 4723148, 4753792, 4753794, 4777999, 4778000, 4906228, 4 |
|---|
Status history