RecallRadar

FDA Device recall Z-2809-2026

EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model N…

On 2026-08-05 the FDA classified a Class II recall of EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model N… by Medtronic Perfusion Systems, citing pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforat….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2809-2026
ClassificationClass II
Statusongoing
Initiated2026-06-24
Published2026-08-05
Last updated2026-09-13 11:58 UTC
CompanyMedtronic Perfusion Systems
DistributionUS

Product

EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722

Reason

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

Identifiers

code_info
UDI-DI: 20199150023312, Lot Numbers: 0232657469 0232670800 0232744945 0232798410 0232862769 0232985251 0233116924 02331…UDI-DI: 20199150023312, Lot Numbers: 0232657469 0232670800 0232744945 0232798410 0232862769 0232985251 0233116924 0233116931; UDI-DI: 00199150023318, Lot Numbers: 0232744945

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2809-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.