FDA Device recall Z-2807-2015
EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used with an Olympus video system center, li…
- FDA Device
- Class II
- terminated
- Published 2015-09-30
On 2015-09-30 the FDA classified a Class II recall of EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used with an Olympus video system center, li… by Olympus America, citing new and validated reprocessing procedures have been issued for the Olympus TJF-Q180V duodenoscope ("TJF-Q180V"), consisting of revised manual cleaning and high level disinfection….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2807-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-03-26 |
| Published | 2015-09-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Olympus America |
| Distribution | US |
Product
EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitors, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
Reason
New and validated reprocessing procedures have been issued for the Olympus TJF-Q180V duodenoscope ("TJF-Q180V"), consisting of revised manual cleaning and high level disinfection procedures.
Identifiers
| code_info | Model: TJF-Q180V, all sold units |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2807-2015%22