FDA Device recall Z-2804-2015
The Leica ASP6025 Tissue Processor
- FDA Device
- Class II
- terminated
- Published 2015-09-30
On 2015-09-30 the FDA classified a Class II recall of The Leica ASP6025 Tissue Processor by Leica Microsystems, citing the ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching le….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2804-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-27 |
| Published | 2015-09-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Leica Microsystems |
| Distribution | AR, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, WV |
Product
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
Reason
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
Identifiers
| code_info | Product Code: Leica ASP6025 Serial Numbers: 163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743, 745, 747, 749,…Product Code: Leica ASP6025 Serial Numbers: 163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743, 745, 747, 749, 751, 753, 755, 757, 759, 761, 763, 765, 767, 769, 771, 773, 775, 777, 779, 781, 783, 785, 787, 789, 791, 793, 795, 797, 799, 801, 803, 805, 807, 809, 811. |
|---|
Status history