FDA Device recall Z-2804-2015
The Leica ASP6025 Tissue Processor
- FDA Device
- Class II
- terminated
- Published 2015-09-30
On 2015-09-30 the FDA classified a Class II recall of The Leica ASP6025 Tissue Processor by Leica Microsystems, citing the ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching le….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2804-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-27 |
| Published | 2015-09-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Leica Microsystems |
| Distribution | AR, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, WV |
Product
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
Reason
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
Identifiers
| code_info | Product Code: Leica ASP6025 Serial Numbers: 163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743, 745, 747, 749,…Product Code: Leica ASP6025 Serial Numbers: 163, 169, 200-729, 731, 733, 735, 737, 739, 741, 743, 745, 747, 749, 751, 753, 755, 757, 759, 761, 763, 765, 767, 769, 771, 773, 775, 777, 779, 781, 783, 785, 787, 789, 791, 793, 795, 797, 799, 801, 803, 805, 807, 809, 811. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Leica Microsystems
- Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system
FDA Device · Class II · 2022-12-21 · Leica Microsystems - Leica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope system
FDA Device · Class II · 2022-12-21 · Leica Microsystems - Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard)
FDA Device · Class II · 2022-12-21 · Leica Microsystems - Leica Surgical Operating Microscopes, Arveo, Part Number 10448999
FDA Device · Class II · 2022-10-05 · Leica Microsystems - Leica Surgical Operating Microscopes, M530 OHX, Part Number 10448737
FDA Device · Class II · 2022-10-05 · Leica Microsystems
All Leica Microsystems recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2804-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.