FDA Device recall Z-2800-2020
uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to a…
- FDA Device
- Class II
- terminated
- Published 2020-08-19
On 2020-08-19 the FDA classified a Class II recall of uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to a… by Shanghai United Imaging Healthcare, citing 1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift c….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2800-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-07-16 |
| Published | 2020-08-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Shanghai United Imaging Healthcare |
| Distribution | US |
Product
uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Reason
1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)
Identifiers
| code_info | Serial No.s 200017, 200023, 200024 and 200036. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2800-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.