On 2015-09-30 the FDA classified a Class II recall of PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-… by Zien Medical Technologies, citing cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause ana….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2800-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-08-18 |
| Published | 2015-09-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zien Medical Technologies |
| Distribution | US |
Product
PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15
Reason
Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.
Identifiers
| code_info | Product Code PB-M-10, Lots 140901J-A, 150228PB10, 150509PB10, 150510PB10; Product Code PB-M-11P, Lots 140901J-B, 150227…Product Code PB-M-10, Lots 140901J-A, 150228PB10, 150509PB10, 150510PB10; Product Code PB-M-11P, Lots 140901J-B, 150227PB11, 150318PB11, 150509PB11, 150510PB11; Product Code PB-M-15, Lots 140901J-C, 150226PB15, 150326PB15, 150511PB15, 150512PB15, 150529PB15. |
|---|