FDA Device recall Z-2793-2018
NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessor…
- FDA Device
- Class II
- terminated
- Published 2018-08-22
On 2018-08-22 the FDA classified a Class II recall of NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessor… by Nxstage Medical, citing there is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2793-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-06-11 |
| Published | 2018-08-22 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Nxstage Medical |
| Distribution | US |
Product
NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessory to the NxStage System One used to warm therapy fluids prior to administration.
Reason
There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.
Identifiers
| code_info | All units are affected. As of 6/11/18, the last serial number manufactured was W35452. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.