FDA Device recall Z-2793-2018

NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessor…

On 2018-08-22 the FDA classified a Class II recall of NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessor… by Nxstage Medical, citing there is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2793-2018
ClassificationClass II
Statusterminated
Initiated2018-06-11
Published2018-08-22
Last updated2026-09-13 11:54 UTC
CompanyNxstage Medical
DistributionUS

Product

NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessory to the NxStage System One used to warm therapy fluids prior to administration.

Reason

There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.

Identifiers

code_infoAll units are affected. As of 6/11/18, the last serial number manufactured was W35452.