FDA Device recall Z-2791-2017
STERILE SAMPLE SORIN GROUP, SMART PERFUSION PACK, NEONATE/PEDIATRIC PACK, REF 627127703, STERILE EO, Rx Only, 1 EA
- FDA Device
- Class II
- terminated
- Published 2017-08-09
On 2017-08-09 the FDA classified a Class II recall of STERILE SAMPLE SORIN GROUP, SMART PERFUSION PACK, NEONATE/PEDIATRIC PACK, REF 627127703, STERILE EO, Rx Only, 1 EA by Sorin Group Usa, citing medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter).….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2791-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-06-29 |
| Published | 2017-08-09 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Sorin Group Usa |
| Distribution | US |
Product
STERILE SAMPLE SORIN GROUP, SMART PERFUSION PACK, NEONATE/PEDIATRIC PACK, REF 627127703, STERILE EO, Rx Only, 1 EA
Reason
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
Identifiers
| code_info | Lot Number 1709500117 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2791-2017%22