FDA Device recall Z-2788-2016
AccuLIF PL 10-16mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with…
- FDA Device
- Class II
- terminated
- Published 2016-09-21
On 2016-09-21 the FDA classified a Class II recall of AccuLIF PL 10-16mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with… by Howmedica Osteonics, citing since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2788-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-07-26 |
| Published | 2016-09-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Howmedica Osteonics |
| Distribution | US |
Product
AccuLIF PL 10-16mm x 11mm x 27mm x 8 Cage, Rx only, Sterile R, Legal The AccuLIF PL Cages are indicated for intervertebral body fusion with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Additionally, the AccuLIF PL Cages can be used as in adjunct to fusion in patients diagnosed with degenerative scoliosis. The AccuLIF Pl Cages are always to be used with supplemental internal spinal fixation. Additionally, the AccuLIF PL Cages are to be used with autograft and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion
Reason
Since Stryker acquired the AccuLIF product in March of 2014, there have been Product Inquires (PIs) related to the PL implant confirming to have a reduction of height > 1 mm. The current occurrence rate for reported post-operative reduction of cage height (> 1mm) of AccuLIF PL Expandable Cages is approximately 0.47% and has demonstrated a slight upward trend since September 2015.
Identifiers
| code_info | Catalog #400010 Lot #'s 01021403, 01121508, 01121509, 01271515, 01271516, 02021512, 02021513, 02021514, 02021515, 0202…Catalog #400010 Lot #'s 01021403, 01121508, 01121509, 01271515, 01271516, 02021512, 02021513, 02021514, 02021515, 02021602, 03221310, 04141508, 04141509, 04141510, 04141511, 04141512, 04251312, 06101405, 06101406, 06101407, 06251311, 07011406, 07201510, 07201511, 07201512, 08141303, 08141313, 09251504, 09281512, 10081301, 10081302, 10201502, 10201503, 10301504, 11121514, 11121518, 11121521, 12151511 and 12151517 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2788-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.