FDA Device recall Z-2786-2017
SORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EA
- FDA Device
- Class II
- terminated
- Published 2017-08-09
On 2017-08-09 the FDA classified a Class II recall of SORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EA by Sorin Group Usa, citing medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter).….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2786-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-06-29 |
| Published | 2017-08-09 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Sorin Group Usa |
| Distribution | US |
Product
SORIN GROUP, SMART PERFUSION PACK, INSPIRE 6FS, REF 044008204, STERILE EO, Rx Only, 1 EA
Reason
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
Identifiers
| code_info | Lot Numbers: 1626400126, 1628100032, 1631900034, 1632800034, 1703300043, 1707400063, 1708000042, 1710000040, 1716500018 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2786-2017%22