FDA Device recall Z-2785-2015

Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiat…

On 2015-09-23 the FDA classified a Class II recall of Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiat… by Raysearch Laboratories Ab, citing an error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose diffe….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2785-2015
ClassificationClass II
Statusterminated
Initiated2015-05-29
Published2015-09-23
Last updated2026-09-13 11:53 UTC
CompanyRaysearch Laboratories Ab
DistributionUS

Product

Radiation Therapy Treatment Planning System software RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

Reason

An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images

Identifiers

code_infosoftware build numbers are 4.7.0.15 or 4.7.1.10

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.