FDA Device recall Z-2777-2016

Integra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra¿ Dermal Regeneration Template, (Integra template…

On 2016-09-21 the FDA classified a Class II recall of Integra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra¿ Dermal Regeneration Template, (Integra template… by Integra Lifesciences, citing integra's post QA release review of historical product release test results for Meshed IDRT products identified the Peel Strength test average result was incorrectly calculated fo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2777-2016
ClassificationClass II
Statusterminated
Initiated2016-03-11
Published2016-09-21
Last updated2026-09-13 11:54 UTC
CompanyIntegra Lifesciences
DistributionUS

Product

Integra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra¿ Dermal Regeneration Template, (Integra template) is a bilayer membrane system for skin replacement. The dermal replacement layer is made of a porous matrix of fibers of cross-linked bovine tendon collagen and glycosaminoglycan (chondroitin-6-sulfate) that is manufactured with a controlled porosity and defined degradation rate. The epidermal substitute layer is made of a thin polysiloxane (silicone) layer to control moisture loss from the wound. Integra template is indicated for the postexcisional treatment of life-threatening full-thickness or deep partial-thickness thermal injuries where sufficient autograft is not available at the time of excision or not desirable due to the physiological condition of the patient. Integra template is also indicated for the repair of scar contractures when other therapies have failed or when donor sites for repair are not sufficient or desirable due to the physiological condition of the patient Manufactured by: Integra LifeSciences Corporation 311 Enterprise Drive, Plainsboro, NJ 08536 877-444-1122 USA n 609-936-5400 outside USA 866-800-7742 fax

Reason

Integra's post QA release review of historical product release test results for Meshed IDRT products identified the Peel Strength test average result was incorrectly calculated for a single Lot (Lot 105A00324750).

Identifiers

code_infoCatalogue No. MIDRT 8101 Lot No. 105A00324750

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.