FDA Device recall Z-2776-2016
CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO…
- FDA Device
- Class II
- terminated
- Published 2016-09-21
On 2016-09-21 the FDA classified a Class II recall of CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO… by Biosense Webster, citing biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosens….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2776-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-08-15 |
| Published | 2016-09-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Biosense Webster |
| Distribution | n/a |
Product
CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO 3 EP Navigation System is intended to be used in catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and catheter location during the procedure.
Reason
Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).
Identifiers
| code_info | V 3.2: 2258, 10284, 10332, 10345, 11640, 11642, 60011, 60012, 60606, 60611, 60613, 60618, 60621, 60625, 60634, 60644, 6…V 3.2: 2258, 10284, 10332, 10345, 11640, 11642, 60011, 60012, 60606, 60611, 60613, 60618, 60621, 60625, 60634, 60644, 60651, 60654, 60657, 60658, 60660, 60661, 60664, 60671, 60673, 60674, 60680, 60682, 60683, 61000, 61002, 61005, 61010, 61014, 61038, 61039, 61041, 61045, 61046, 61050, 61055, 61060, 61062, 61064, 61071, 61501, 66114, 66115, 66117, 66118, 66119, 29306, 13173, 13213, 11567, 14528 V 4.3: 61053, 61054, 61059, 11701, 12664 V 2.3: 14218, 13208, 50203, 50531, 14718, 50573 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2776-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.