FDA Device recall Z-2775-2018
Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-…
- FDA Device
- Class II
- terminated
- Published 2018-09-26
On 2018-09-26 the FDA classified a Class II recall of Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-… by Degania Silicone, citing the catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2775-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-06-21 |
| Published | 2018-09-26 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Degania Silicone |
| Distribution | US |
Product
Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Reason
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Identifiers
| code_info | S17003407 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2775-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.