RecallRadar

FDA Device recall Z-2774-2017

STOCKERT Heater-Cooler System 3T, 127V/60 Hz

On 2017-08-02 the FDA classified a Class II recall of STOCKERT Heater-Cooler System 3T, 127V/60 Hz by Sorin Group Usa, citing plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceed….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2774-2017
ClassificationClass II
Statusterminated
Initiated2017-06-30
Published2017-08-02
Last updated2026-09-13 11:54 UTC
CompanySorin Group Usa
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

STOCKERT Heater-Cooler System 3T, 127V/60 Hz

Reason

Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.

Identifiers

code_infoItem No. 16-02-83; Serial No. 16S11455-16S16509