Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet a…
- FDA Device
- Class II
- ongoing
- Published 2020-08-12
On 2020-08-12 the FDA classified a Class II recall of Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet a… by Covidien, citing this voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved d….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2769-2020 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-06-26 |
| Published | 2020-08-12 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Covidien |
| Distribution | n/a |
Product
Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.
Reason
This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
Identifiers
| code_info | Model # DSA-135-01 UDI: 10884521663527 Lot #: F2510371X, F2510861X, F2512341X, F2512383X, F2512493X, F2513453X |
|---|