RecallRadar

FDA Device recall Z-2757-2026

FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004…

On 2026-07-29 the FDA classified a Class II recall of FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004… by Boston Scientific, citing boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orien….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2757-2026
ClassificationClass II
Statusongoing
Initiated2026-06-11
Published2026-07-29
Last updated2026-09-13 11:58 UTC
CompanyBoston Scientific
DistributionUS

Product

FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01

Reason

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Identifiers

code_info
UDI-DI: 00191506042516; Lots: CL12917, CL13004, CL13017, CL13046, CL13060, CL13080, CL13103, CL13122, CL13123, CL13142,…UDI-DI: 00191506042516; Lots: CL12917, CL13004, CL13017, CL13046, CL13060, CL13080, CL13103, CL13122, CL13123, CL13142, CL13160, CL13196, CL13212, CL13219, CL13231, CL13260, CL13291, CL13340, CL13375, CL13401, CL13422, CL13468, CL13519, CL13573, CL13578, CL13614, CL13646, CL13670, CL13671, CL13708, CL13719, CL13722, CL13732, CL13760, CL13801, CL13802, CL13820, CL13833, CL13856, CL13910, CL13911, CL13925, CL13940, CL13971, CL13972, CL13988, CL14025, CL14039, CL14050, CL14051, CL14076, CL14077, CL14119, CL14120, CL14135, CL14137, CL14171, CL14187, CL14226, CL14249, CL14286, CL14287, CL14345, CL14346, CL14375, CL14376, CL14396, CL14397, CL14439, CL14488, CL14500, CL14501, CL14518, CL14519, CL14581, CL14582, CL14633, CL14635, CL14670, CL14697, CL14711, CL14749, CL14750, CL14769, CL14770, CL14813, CL14814, CL14854, CL14855, CL14870, CL14871, CL14928, CL14929, CL14933, CL14934, CL14935, CL14968, CL14969, CL15005, CL15009, CL15067, CL15107, CL15108, CL15109, CL15128, CL15129, CL15191, CL15192

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2757-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.