FDA Device recall Z-2750-2018
CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The CDI System 500 provides continuous, on-line monitoring…
- FDA Device
- Class II
- terminated
- Published 2018-08-22
On 2018-08-22 the FDA classified a Class II recall of CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The CDI System 500 provides continuous, on-line monitoring… by Terumo Cardiovascular Systems, citing H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automate….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2750-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-06-26 |
| Published | 2018-08-22 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37¿C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.
Reason
H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.
Identifiers
| code_info | H011684 H011685 H011686 H011687…H011684 H011685 H011686 H011687 H011688 H011689 H011691 H011692 H011693 H011694 H011695 H011696 H011697 H011699 H011700 H011701 H011702 H011704 H011705 H011706 H011707 H011708 H011709 H011710 H011711 H011712 H011713 H011714 H011715 H011717 H011718 H011719 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.