FDA Device recall Z-2746-2026

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AI…

On 2026-07-22 the FDA classified a Class II recall of ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AI… by Tosoh Bioscience, citing the Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevate….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2746-2026
ClassificationClass II
Statusongoing
Initiated2026-06-12
Published2026-07-22
Last updated2026-09-13 11:58 UTC
CompanyTosoh Bioscience
DistributionUS

Product

ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204

Reason

The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.

Identifiers

code_infoUDI-DI: 14560189210032; Lot Numbers: FX12404, FY12407, FY12406, FZ12408

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.