G8 Automated HPLC Analyzer: HLC-723G8-ST, 021560

On 2018-08-22 the FDA classified a Class II recall of G8 Automated HPLC Analyzer: HLC-723G8-ST, 021560 by Tosoh Bioscience, citing hbAE is known to interfere with the HbA1c assay on the current version of software, Ver. 5.23. Customers should exercise caution when reviewing chromatograms and ensure that Flag….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2745-2018
ClassificationClass II
Statusterminated
Initiated2018-06-06
Published2018-08-22
Last updated2026-09-13 11:54 UTC
CompanyTosoh Bioscience
DistributionUS

Product

G8 Automated HPLC Analyzer: HLC-723G8-ST, 021560; and HLC-723G8-LA, 021674 Product Usage: The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in whole blood specimens. This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.

Reason

HbAE is known to interfere with the HbA1c assay on the current version of software, Ver. 5.23. Customers should exercise caution when reviewing chromatograms and ensure that Flag 43 is enabled on their device to avoid reporting invalid test results in the presence of HbAE.

Identifiers

code_infoAll active analyzers with Version 5.23 Software

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.