RecallRadar

FDA Device recall Z-2744-2026

Photera 600 3-D, phototherapy device

On 2026-07-22 the FDA classified a Class II recall of Photera 600 3-D, phototherapy device by Daavlin Distributing, citing calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2744-2026
ClassificationClass II
Statusongoing
Initiated2026-03-05
Published2026-07-22
Last updated2026-09-13 11:58 UTC
CompanyDaavlin Distributing
DistributionUS

Product

Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Reason

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Identifiers

code_info
Serial Numbers: 6-09278NB8CL, 6-09271NB8CL, 6-09055NB8CL, 6-09113NB8CL, 6-09204NB8CL, 6-09207NB8CL, 6-09116NB8CL, 6-09…Serial Numbers: 6-09278NB8CL, 6-09271NB8CL, 6-09055NB8CL, 6-09113NB8CL, 6-09204NB8CL, 6-09207NB8CL, 6-09116NB8CL, 6-09263NB8CL, 6-09272NB8CL, 6-09118NB8CL, 6-09275NB8CL, 6-09264NB8CL, 6-09060NB8CL, 6-09122NB8CL, 6-09216NB8CL, 6-09125NB8CL, 6-09353NB8CL, 6-09346NB8CL, 6-09128NB8CL, 6-09129NB8CL, 6-09133NB8CL, 6-09134NB8CL, 6-09266NB8CL, 6-09215NB8CL, 6-09355NB8CL, 6-09137NB8CL, 6-09359NB8CL, 6-09343NB8CL, 6-09140NB8CL, 6-09142NB8CL, 6-09143NB8CL, 6-09273NB8CL, 6-09065NB8CL, 6-09144NB8CL, 6-09237NB8CL, 6-09225NB8CL, 6-09145NB8CL, 6-09147NB8CL, 6-09148NB8CL, 6-09378NB8CL, 6-09252NB8CL, 6-09051NB8CL, 6-09260NB8CL, 6-09250NB8CL, 6-09149NB8CL, 6-09238NB8CL, 6-09212NB8CL, 6-09277NB8CL, 6-09048NB8CL, 6-09274NB8CL, 6-09259NB8CL, 6-09224NB8CL, 6-09057NB8CL, 6-09436NB12CL, 6-09438NB8CL, 6-09434NB12CL, 6-09270NB8CL, 6-09115NB8CL, 6-09111NB8CL, 6-09435NB12CL, 6-09124NB8CL, 6-09332NB12CL, 6-09223NB8CL, 6-09219NB8CL, 6-09234NB8CL, 6-09063NB8CL, 6-09232NB8CL, 6-09479NB12CL, 6-09373NB8CL, 6-09127NB8CL