RecallRadar

FDA Device recall Z-2738-2015

2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plati…

On 2015-09-16 the FDA classified a Class II recall of 2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plati… by Trilliant Surgical, citing gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the scre….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2738-2015
ClassificationClass II
Statusterminated
Initiated2015-07-31
Published2015-09-16
Last updated2026-09-13 11:53 UTC
CompanyTrilliant Surgical
DistributionAR, AZ, CA, FL, GA, IA, IL, IN, MN, MO, NJ, NY, OH, OK, SC, SD, TX, VA, WI

Product

2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.

Reason

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Identifiers

code_infoTSL002605

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2738-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.