RecallRadar

FDA Device recall Z-2736-2020

Alaris Pump Module Model 8100

On 2020-08-12 the FDA classified a Class I recall of Alaris Pump Module Model 8100 by Carefusion 303, citing broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dro….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2736-2020
ClassificationClass I
Statusterminated
Initiated2020-06-30
Published2020-08-12
Last updated2026-09-13 11:55 UTC
CompanyCarefusion 303
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

Alaris Pump Module Model 8100

Reason

Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.

Identifiers

code_infoAll serial numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.