FDA Device recall Z-2735-2017
Agfa Healthcare NX 3.0.8950 Imaging Processing Software
- FDA Device
- Class II
- terminated
- Published 2017-07-26
On 2017-07-26 the FDA classified a Class II recall of Agfa Healthcare NX 3.0.8950 Imaging Processing Software by Agfa Healthcare, citing A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong ima….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2735-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-12-21 |
| Published | 2017-07-26 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Agfa Healthcare |
| Distribution | AL, AR, AZ, CA, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WV |
Product
Agfa Healthcare NX 3.0.8950 Imaging Processing Software
Reason
A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.
Identifiers
| code_info | 2UA15205XC, 2UA161JNH, A5411002007, A5411001003, A5411001002, 5CG5164040, 2UA5481BVY, 5CG5171X81, A5411002026, A5411002…2UA15205XC, 2UA161JNH, A5411002007, A5411001003, A5411001002, 5CG5164040, 2UA5481BVY, 5CG5171X81, A5411002026, A5411002025, A5410000039, 2UA62524PS, 2UA62524Q2, 2UA62524RK, A5411002031, 2UA61218G4, 5CG61266X6, 2UA61218GG, 2UA63825HQ, 2UA642306H, 2UA4161JMY, 2UA52524Y5, 2UA62524RJ, 2UA62524QV, 2UA62524QB, 2UA62524RJ, 2UA62524NS, 2UA62524PJ, 2UA62524Q9, A5411000155, A5411000151, 2UA62524PH, 2UA62524QL, 2UA62524QT, A5411000153, 2UA62524Q0, 5CG6126B9J, 5CG61266YW, 5CG61266VP, 5CG61268BG, 5CG612673N, 2UA6243GNY, 2UA62424R8, 2UA62524P4, 2UA62524RY, 2UA6243GPT, 2UA62524Q6, 2UA62524PR, A5411000152, 2UA63825HR, 2UA31000Q6, 2UA31000Q8, 2UA31000R5, 2UA31000QR, 2UA319120W, 2UA3191211, 2UA31000PS, 2UA31000QX, 2UA31000Q0, 2UA31000PX, A5411000346, A5411000232, A541000199, A5411000210, A5411000209, 2UA62524P8, 5CG61266R5, 5CG5295760, 5CG5285W67, 5C5285W7P, 5CG5285W9Y, 5CG5285W5R, 5CG528W7B, A5411001163, A5411000468, 2UA52524XM, 2UA5481BTP, A5411000145, 2UA25009HM, A5411001064, A5411000159, 2UA3380VMW, |
|---|