FDA Device recall Z-2732-2017

Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer w…

On 2017-07-19 the FDA classified a Class II recall of Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer w… by Nova Biomedical, citing software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2732-2017
ClassificationClass II
Statusterminated
Initiated2017-06-19
Published2017-07-19
Last updated2026-09-13 11:54 UTC
CompanyNova Biomedical
DistributionUS

Product

Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime CCS Analyzer PN 53656 Stat Profile Prime CCS Analyzer with Scanner PN 53657 Stat Profile Prime CCS Comp Analyzer with Scanner PN 52857 Stat Profile Prime CCS Comp Analyzer Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer

Reason

Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.

Identifiers

code_infoAll current software versions

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2732-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.