FDA Device recall Z-2730-2017

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Gluco…

On 2017-07-19 the FDA classified a Class II recall of Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Gluco… by Beckman Coulter, citing the firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2730-2017
ClassificationClass II
Statusterminated
Initiated2017-05-15
Published2017-07-19
Last updated2026-09-13 11:54 UTC
CompanyBeckman Coulter
DistributionUS

Product

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Reason

The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

Identifiers

code_infoLot No. 72645074, Exp. Date 30 Apr 2018

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.