FDA Device recall Z-2728-2024
Stryker iBur 4.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shap…
- FDA Device
- Class II
- completed
- Published 2024-09-04
On 2024-09-04 the FDA classified a Class II recall of Stryker iBur 4.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shap… by Stryker, citing potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the po….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2728-2024 |
| Classification | Class II |
| Status | completed |
| Initiated | 2024-07-24 |
| Published | 2024-09-04 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Stryker |
| Distribution | US |
Product
Stryker iBur 4.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-040DC
Reason
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Identifiers
| code_info | GTIN: 07613327501155 Lot Numbers: 22248017 22263017 22334017 22348017 22348027 23016027 23095027 23128017 23163017 2317…GTIN: 07613327501155 Lot Numbers: 22248017 22263017 22334017 22348017 22348027 23016027 23095027 23128017 23163017 23179017 23205017 23205027 23205037 23221017 23311017 24004017 24004027 24004037 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: completed.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2728-2024%22