RecallRadar

FDA Device recall Z-2728-2015

Transonic Flow Probe, 4 mm, Catalog Number HQD4FMC -- Product Usage: to measure flow intra-operatively

On 2015-09-09 the FDA classified a Class II recall of Transonic Flow Probe, 4 mm, Catalog Number HQD4FMC -- Product Usage: to measure flow intra-operatively by Transonic Systems, citing software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow met….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2728-2015
ClassificationClass II
Statusterminated
Initiated2015-07-27
Published2015-09-09
Last updated2026-09-13 11:53 UTC
CompanyTransonic Systems
DistributionUS

Product

Transonic Flow Probe, 4 mm, Catalog Number HQD4FMC -- Product Usage: to measure flow intra-operatively.

Reason

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

Identifiers

code_info
Serial Numbers: HQD4FMC3504, HQD4FMC3505, HQD4FMC3506, HQD4FMC3507, HQD4FMC3508, HQD4FMC3509, HQD4FMC3510, HQD4FMC3511…Serial Numbers: HQD4FMC3504, HQD4FMC3505, HQD4FMC3506, HQD4FMC3507, HQD4FMC3508, HQD4FMC3509, HQD4FMC3510, HQD4FMC3511, HQD4FMC3512, HQD4FMC3513, HQD4FMC3514, HQD4FMC3515, HQD4FMC3516, HQD4FMC3517, HQD4FMC3518, HQD4FMC3519, HQD4FMC3520

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2728-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.