FDA Device recall Z-2727-2024
Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shap…
- FDA Device
- Class II
- completed
- Published 2024-09-04
On 2024-09-04 the FDA classified a Class II recall of Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shap… by Stryker, citing potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the po….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2727-2024 |
| Classification | Class II |
| Status | completed |
| Initiated | 2024-07-24 |
| Published | 2024-09-04 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Stryker |
| Distribution | US |
Product
Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-030DC
Reason
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Identifiers
| code_info | GTIN: 07613327501209 Lot Numbers: 22230017 22333017 22342017 23011017 23093017 23172017 23188017 23221017 23311017 24004027 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: completed.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2727-2024%22