FDA Device recall Z-2727-2015
Transonic Flow Probe, 3 mm, Catalog Number HQD3FMC -- Product Usage: to measure flow intra-operatively
- FDA Device
- Class II
- terminated
- Published 2015-09-09
On 2015-09-09 the FDA classified a Class II recall of Transonic Flow Probe, 3 mm, Catalog Number HQD3FMC -- Product Usage: to measure flow intra-operatively by Transonic Systems, citing software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow met….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2727-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-27 |
| Published | 2015-09-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Transonic Systems |
| Distribution | US |
Product
Transonic Flow Probe, 3 mm, Catalog Number HQD3FMC -- Product Usage: to measure flow intra-operatively.
Reason
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
Identifiers
| code_info | Serial Numbers: HQD3FMC3763, HQD3FMC3764, HQD3FMC3765, HQD3FMC3766, HQD3FMC3767, HQD3FMC3768, HQD3FMC3769, HQD3FMC3770…Serial Numbers: HQD3FMC3763, HQD3FMC3764, HQD3FMC3765, HQD3FMC3766, HQD3FMC3767, HQD3FMC3768, HQD3FMC3769, HQD3FMC3770, HQD3FMC3771, HQD3FMC3772, HQD3FMC3773, HQD3FMC3774, HQD3FMC3775, HQD3FMC3776, HQD3FMC3777, HQD3FMC3778, HQD3FMC3779, HQD3FMC3780, HQD3FMC3781, HQD3FMC3782, HQD3FMC3783, HQD3FMC3784, HQD3FMC3785, HQD3FMC3786, HQD3FMC3787, HQD3FMC3788, HQD3FMC3789, HQD3FMC3790, HQD3FMC3791, HQD3FMC3792, HQD3FMC3793, HQD3FMC3794, HQD3FMC3795, HQD3FMC3796, HQD3FMC3797, HQD3FMC3798, HQD3FMC3799, HQD3FMC3800, HQD3FMC3801, HQD3FMC3802, HQD3FMC3803, HQD3FMC3804, HQD3FMC3805, HQD3FMC3806, HQD3FMC3807, HQD3FMC3814, HQD3FMC3815 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2727-2015%22